عنوان
|
Voltammetric determination of dopamine using modified glassy carbon electrode by electrografting of catechol
|
نوع پژوهش
|
مقاله چاپشده در مجلات علمی
|
کلیدواژهها
|
electrografted glassy carbon electrode; dopamine; catechol; cyclic voltammetry; electrochemical sensor.
|
چکیده
|
By means of cyclic voltammetry as a diagnostic technique, the present study tried to describe the behaviour of dopamine (DP) at the bare and catechol electrografted glassy carbon electrode (CA/GCE). The results indicated that the CA/GCE exhibits high electrocatalytic activity toward DP. The DP cycling of the CA/GCE in DP showed two linear calibration ranges between 5.0–100 and 100–750 μM with a detection limit of 0.86 μM. For 6 DP replicate measurements (50 μM); the relative standard deviation (RSD) was 2.7 %. This method was successfully used for the determination of DP in serum samples.
|
پژوهشگران
|
زینب پورقبادی (نفر اول)، داود نعمت الهی (نفر دوم)
|