This study aimed to present a new method combined of needle trap devices (NTDs) packed with the Zn3(BTC)2 metal-organic framework (Zn-MOF) as a sorbent, for the sampling and analysis of polycylic aromatic hydrocarbons (PAHs) in the air for the frst time. The electrochemical method was used to synthesize Zn-MOF sorbent packed inside 22 gauge needle trap. This method was investigated in the laboratory under different condition and applied in the feld to determination of PAHs in air. In order to make different concentrations of PAHs in air, a glass chamber under temperature of 120 °C was used. The optimization of the desorption conditions and breakthrough volume was accomplished using the response surface method and Design-Expert software (version 7). According to the results, the best desorption temperature was obtained as 379 °C in 9 min, the limit of detection and limit of quantifcation for the analytes under investigation were within the ranges of 0.011–0.021 and 0.03–0.07 mg/m3, respectively. Moreover, repeatability and reproducibility of the method were estimated in the ranges of 3.6–9.9 and 5.3–24.1, respectively. By storing the needles at 4 °C, no signifcant reduction was observed in the amount of analytes after 60 days (P > 0.05). It could be concluded that NTD: Zn-MOF is a reliable and high collection efciency method had a suitable performance in the sampling of PAHs, compared to the National Institute for Occupational Safety and Health (NIOSH)-5515 method.